RADWAG MA 200/1.X2.A
Description
MA X2.A moisture analysers are hi-tech measuring instruments intended for quick determination of relative moisture content,
dry mass content and other parameters in samples of different substances.
MA.X2.A series is equipped with an innovative system: the drying chamber can be opened and closed automatically using a button or proximity sensors.
Such a solution allows:
- Maintaining the moisture analyser clean – the operator does not touch the moisture analyser’s housing.
- Removing any shocks caused by manual closing of the chamber – the chamber closes automatically and always with the same intensity.
- Achieving high repeatability of opening and closing the drying chamber.
- Enhancing drying process automation.
- Enhancing safety of operation by eliminating contact with hot components of drying chamber.
- Improving operation ergonomics.
MA.X2.A series features a user-friendly colour touchscreen. Free configuration of the control panel can be done by means of hotkeys, customizable information fields and programmable labels. Drying process can be carried out for any temperature and – thanks to databases – particular drying process may be set for a specific product.
Physical characteristics
Leveling system: | Manual |
Display: | 5” graphic color touchscreen |
Weighing chamber: | Automatic |
Maximum sample height: | 20 mm |
Weighing pan dimensions: | ø90, h= 8 mm |
Packaging dimensions: | 476×381×346 mm |
Net weight: | 5.2 kg |
Gross weight: | 6.5 kg |
Construction: | |
Protection class: | IP 43 |
Features of use: | |
Display modes: | %MC, %DC, %ATRO MC, %ATRO DC, g, g/kg MC, g/kg DC, -%MC |
Drying method: | 4 drying profiles (standard, fast, step, mild) |
Finish mode: | 4 modes (automatic, manual, time-defined, user-defined) |
Additional functions: | sample traceability |
Communication interface: | |
Communication interface: | RS232¹, USB-A, USB-B, Ethernet, Wi-Fi® |
Electrical parameters: | |
Power supply: | 100V – 120V AC 50/60Hz or 200V-240V AC 50/60Hz |
Power consumption: | 6 W |
Heating module power: | 450 W |
Environmental conditions: | |
Operating temperature: | +10 — +40 °C |
SOP – metrological tests: | |
Mass indictation control: | External mass standard 50 g, 100 g, 200 g (depending on the type of moisture analyzer) |
Internal adjustment: | For .IC models only, GLP report |
External adjustment: | For all models, GLP report |
Drying temperature control: | One- or two-point, depending on the drying profile (equipment: GT105k-12/Z Control Thermometer) |
Moisture content indication control: | Certified MV Test Set |
Adjustment: | |
Internal / External adjustment: | Depending on the type of moisture analyzer, GLP report |
Drying temperature adjustment: | Three-point adjustment at defined temperatures, programmable values (equipment: GT105k-12/Z Control Thermometer) |